TB 9-6625-114-24
Table 2. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
Range: 2 kHz @ 800 mV p-p;
Agilent, Model 33250A (33250A)
1 kHz @ 1.2 V p-p, 1.4 V offset
Accuracy: calibrated
AC:
OSCILLOSOPE
Fluke, Model 5820A-5C-GHZ
Range: 0.02 V p-p to 100 V p-p
CALIBRATOR
(5820A-5C-GHz)
Accuracy: 1%
Markers:
Range: 1 ns to 1 ms
Accuracy: 0.01%
Level Sine:
Range: 50 kHz to 100 MHz
Accuracy: 5%
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Complete paragraphs 8 through 22 before attempting adjustments for data tracking.
When indications specified in paragraphs 8 through 22 are not within tolerance, perform
pertinent adjustments in paragraph 23. After adjustments are made, repeat paragraphs as
instructed in paragraph 23. Once TI indications are within tolerance, move to paragraph 24.
e. Unless otherwise specified, all controls and control settings refer to the TI.