Adjustments - Continued

Calibration Procedure For Scopemeter Models 93, 95 And 97 Page Navigation
  13    14    15    16    17  18
img
. . . . .
TB 9-6625-114-24
NOTE
If all connections are correct and the correct probe was used, TI
will display "DC PROBE calibration in progress." After a
few  seconds  TI  will  display  "PROBE  successfully
calibrated" and will beep once.
(69) Press TI CAL STORE soft-key.
(70) Turn TI off.
(71) Press and hold TI LCD key while pressing TI ON/OFF key (when TI switches
on, there will be two beeps). Release TI LCD and ON/OFF keys.
(72) If only TI METER mode adjustments (paragraph 23 b) were performed, perform
paragraphs 15 through 22 and 24. If TI SCOPE and METER mode adjustments (23 a and
b) were performed, perform paragraphs 8 through 22 and 24.
24. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form IAW TB 750-25.
19/(20 blank)