TB 9-6625-114-24
(21) Set calibrator for a 300 V dc output.
(22) Press TI READY soft-key.
(23) Press TI upper select/adjust key ( ) (step 8).
(24) Set calibrator output to STBY and connect calibrator OUTPUT HI and LO to TI
mV (red) and COM (black) inputs observing polarity.
(25) Set calibrator for a 300 mV dc output.
(26) Press TI READY soft-key.
(27) Press TI upper select/adjust key ( ) (step 9).
(28) Set calibrator for a 3 V dc output.
(29) Press TI READY soft-key.
(30) Press TI upper select/adjust key ( ) (step 10).
(31) Disconnect calibrator from TI and short TI mV (red) and COM (black) inputs.
(32) Press TI READY soft-key.
(33) Press TI upper select/adjust key ( ) (step 11).
(34) Set calibrator for a 100 Ÿ output.
(35) Press TI READY soft-key.
(36) Press TI upper select/adjust key ( ) (step 12).
(37) Set calibrator for a 1 kŸ output.
(38) Press TI READY soft-key.
(39) Press TI upper select/adjust key ( ) (step 13).
(40) Set calibrator for a 10 kŸ output.
(41) Press TI READY soft-key.
(42) Press TI upper select/adjust key ( ) (step 14).
(43) Set calibrator for a 100 kŸ output.
(44) Press TI READY soft-key.
(45) Press TI upper select/adjust key ( ) (step 15).
(46) Set calibrator for a 1 MŸ output.
(47) Press TI READY soft-key.
(48) Press TI upper select/adjust key ( ) (step 16).
(49) Set calibrator for a 10 MŸ output.
(50) Press TI READY soft-key.
(51) Press TI upper select/adjust key ( ) (step 17).
(52) Disconnect calibrator from TI and connect TI CHANNEL A to TI
(53) Press TI READY soft-key.
(54) Press TI upper select/adjust key ( ) (step 18).
17