Adjustments - Continued
Adjustments - Continued

Calibration Procedure For Scopemeter Models 93, 95 And 97 Page Navigation
  10    11    12    13    14  15  16    17    18  
img
. . . .
TB 9-6625-114-24
NOTE
If only TI SCOPE mode adjustments are required, perform
steps (78) through (80) below. If TI METER mode adjustment
is also required, disregard (78) through (81) below and continue
with paragraph 23 b.
(78) Press TI CAL STORE soft-key.
(79) Turn TI off.
(80) Press and hold TI LCD key while pressing TI ON/OFF key (when TI switches
on, there will be two beeps). Release TI LCD and ON/OFF keys.
(81) Perform paragraphs 8 through 14 and 24.
b. Meter Mode
NOTE
If performing only TI METER mode adjustments, paragraph
23 a steps (1) through (10) above must first be performed.
(1) Press TI METER soft-key.
(2) Press TI upper select/adjust key ( ) (step 1).
(3) Short TI CHANNEL A BNC input and TI banana connectors.
(4) Press TI READY soft-key.
(5) Press TI upper select/adjust key ( ) (step 2).
(6) Press TI READY soft-key.
(7) Press TI upper select/adjust key ( ) (step 3).
(8) Remove short on CHANNEL A input.
(9) Press TI READY soft-key.
(10) Press TI upper select/adjust key ( ) (step 4).
(11) Connect calibrator OUTPUT HI and LO to TI CHANNEL A input.
(12) Set calibrator for a 300 mV dc output.
(13) Press TI READY soft-key.
(14) Press TI upper select/adjust key ( ) (step 5).
(15) Set calibrator for a 3 V dc output.
(16) Press TI READY soft-key.
(17) Press TI upper select/adjust key ( ) (step 6).
(18) Set calibrator for a 30 V dc output.
(19) Press TI READY soft-key.
(20) Press TI upper select/adjust key ( ) (step 7).
16