TB 9-6625-114-24
(48) Set oscilloscope calibrator for a 1 kHz, 10 V p-p square wave output.
(49) Press TI READY soft-key.
(50) Press TI upper select/adjust key ( ) (step 17).
(51) Set oscilloscope calibrator for a 1 kHz, 100 V p-p square wave output.
(52) Press TI READY soft-key.
(53) Press TI upper select/adjust key ( ) (step 18).
(54) Set oscilloscope calibrator for a 1 kHz, 200 mV p-p square wave output.
(55) Press TI READY soft-key.
(56) Press TI upper select/adjust key ( ) (step 19).
(57) Set oscilloscope calibrator for a 1 kHz, 20 mV p-p square wave output.
(58) Press TI READY soft-key.
(59) Press TI upper select/adjust key ( ) (step 20).
(60) Insert a 50 feedthrough in connection for both TI CHANNEL A and CHANNEL B.
(61) Set oscilloscope calibrator for a 50 kHz, 1 V p-p sine wave output.
(62) Press TI READY soft-key.
(63) Press TI upper select/adjust key ( ) (step 21).
(64) Press TI READY soft-key.
(65) Press TI upper select/adjust key ( ) (step 22).
(66) Press TI READY soft-key.
(67) Press TI upper select/adjust key ( ) (step 23).
(68) Press TI READY soft-key.
(69) Press TI upper select/adjust key ( ) (step 24).
(70) Connect a BNC dual input coupler to oscilloscope calibrator
SOURCE/MEASURE CHAN 1. Connect one end of dual input coupler outputs to TI
CHANNEL A input and connect the other output through a BNC to banana adapter to the
mV (red) and COM (black) inputs observing polarity.
(71) Set oscilloscope calibrator for a 50 kHz, 1 V p-p sine wave output.
(72) Press TI READY soft-key.
(73) Connect a BNC dual input coupler to oscilloscope calibrator
SOURCE/MEASURE CHAN 1. Connect one end of dual input coupler outputs to TI
CHANNEL A through a 50 feedthrough and connect the other end to TI CHANNEL B
through a 50 feedthrough.
(74) Set oscilloscope calibrator for a 1 MHz, 600 mV p-p EDGE output.
(75) Press TI READY soft-key.
(76) Set oscilloscope calibrator output to STBY and press TI SCOPE soft-key.
(77) Disconnect equipment setup.
15