Adjustments - Continued
Final Procedure

Calibration Procedure For Scopemeter Models 93, 95 And 97 Page Navigation
  12    13    14    15    16  17  18  
TB 9-6625-114-24
(55) Leave TI CHANNEL A connected to TI GENERATOR OUT banana connector
and connect 100 Ÿ between TI GENERATOR OUT and COM connectors.
(56) Press TI READY soft-key.
(57) Press TI upper select/adjust key ( ) (step 19).
(58) Disconnect equipment setup and connect red scope probe (p/o TI) to TI
CHANNEL A input. Connect probe tip to TI GENERATOR OUT connector using red
adjust adapter (p/o TI).
(59) Press TI READY soft-key.
NOTE
If all connections are correct and the correct probe was used, TI
will display "DC PROBE calibration in progress". After a
few  seconds  TI  will  display  "PROBE  successfully
calibrated" and will beep once.
(60) Press TI upper select/adjust key ( ) (step 20).
(61) Disconnect equipment setup and connect grey scope probe (p/o TI) to TI
CHANNEL B input. Connect probe tip to TI GENERATOR OUT connector using red
adjust adapter (p/o TI).
(62) Press TI READY soft-key.
NOTE
If all connections are correct and the correct probe was used, TI
will display "DC PROBE calibration in progress." After a
few  seconds  TI  will  display  "PROBE  successfully
calibrated" and will beep once.
(63) Press TI upper select/adjust key ( ) (step 21).
(64) Disconnect equipment setup and connect TI CHANNEL A to TI GENERATOR
OUT banana connector.
(65) Press TI READY soft-key.
NOTE
If all connections are correct and the correct probe was used, TI
will display "DC PROBE calibration in progress." After a
few  seconds  TI  will  display  "PROBE  successfully
calibrated" and will beep once.
(66) Press TI upper select/adjust key ( ) (step 22).
(67) Move connection from TI CHANNEL A input to TI CHANNEL B input.
(68) Press TI READY soft-key.
18