TB 9-6625-114-24
(15) Press TI upper select/adjust key ( ) (step 6).
(16) Connect a BNC dual input coupler to oscilloscope calibrator
SOURCE/MEASURE CHAN 1. Connect one end of dual input coupler outputs to TI
CHANNEL A through a 50 feedthrough and connect the other end to TI CHANNEL B
through a 50 feedthrough.
(17) Set oscilloscope calibrator for a 1 kHz, 300 mV p-p square wave output.
(18) Press TI READY soft-key.
(19) Press TI upper select/adjust key ( ) (step 7).
(20) Set oscilloscope calibrator for a 1 kHz, 3 V p-p square wave output.
(21) Press TI READY soft-key.
(22) Press TI upper select/adjust key ( ) (step 8).
(23) Remove 50 feedthroughs from connection.
(24) Set oscilloscope calibrator for a 1 kHz, 20 V p-p square wave output.
(25) Press TI READY soft-key.
(26) Press TI upper select/adjust key ( ) (step 9).
(27) Set oscilloscope calibrator for a 1 kHz, 50 V p-p square wave output.
(28) Press TI READY soft-key.
(29) Press TI upper select/adjust key ( ) (step 10).
(30) Set oscilloscope calibrator for a 1 kHz, 20 mV p-p square wave output.
(31) Press TI READY soft-key.
(32) Press TI upper select/adjust key ( ) (step 11).
(33) Set oscilloscope calibrator for a 1 kHz, 50 mV p-p square wave output.
(34) Press TI READY soft-key.
(35) Press TI upper select/adjust key ( ) (step 12).
(36) Set oscilloscope calibrator for a 1 kHz, 100 mV p-p square wave output.
(37) Press TI READY soft-key.
(38) Press TI upper select/adjust key ( ) (step 13).
(39) Set oscilloscope calibrator for a 1 kHz, 200 mV p-p square wave output.
(40) Press TI READY soft-key.
(41) Press TI upper select/adjust key ( ) (step 14).
(42) Set oscilloscope calibrator for a 1 kHz, 500 mV p-p square wave output.
(43) Press TI READY soft-key.
(44) Press TI upper select/adjust key ( ) (step 15).
(45) Set oscilloscope calibrator for a 1 kHz, 1 V p-p square wave output.
(46) Press TI READY soft-key.
(47) Press TI upper select/adjust key ( ) (step 16).
14