Common Mode Rejection Ratio

Calibration Procedure For Oscilloscope Model Gos622G Page Navigation
  5    6    7    8    9  10
TB 9-6625-128-24
(9) Set the frequency counter as required to measure frequency.
(10) Verify the frequency counter indicates within 950 to 1050 Hz; if not perform b
below.
(11) Set the oscilloscope calibrator output to standby.
b. Adjustments. No adjustments can be made.
15. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
11/(12 blank)