TB 9-6625-133-40
(7) Allow measurement to stabilize, then record TI indication in table 7.
(8) Calculate TI error using formula below. TI will indicate within the limits of
CTUAL - NOMINAL
ERROR ppm = A
,
,
NOMINAL
Where:
ACTUAL = TI indication
NOMINAL = Test report value of standard resistor
(9) Repeat technique of (6) through (8) above for remaining rows in table 7. Use
table 6 to configure the TI for the resistance mode to be calibrated. TI will indicate within
(10)Ensure that Ω pushbutton has been selected to remove HI Voltage from TI
terminals (see warning above).
(11)Disconnect equipment set-up.
b. Adjustments. Complete all performance checks in the entire procedure noting AS
a. Performance Check
(1) Prepare TI for zeroing by ensuring that the INPUT terminals are open.
(2) Press TI keys as listed in (a) through (i) below.
(a)
CLEAR
(b)
[PWR UP DFLT]
(c)
DCI
(d)
CONFIG
(e)
[FAST] (fast de-selected)
(f)
[RESL] , [7]
(g)
DCI
(h)
INPUT
(i)
[ZERO FUNC] (zero is complete when the BUSY legend goes out)
(3) Populate the true applied column in table 8 with the values obtained during DCI
(4) Connect calibrator OUTPUT HI / LO to TI INPUT A / LO.
(5) Set TI to the first range in table 8.
(6) Set calibrator to the first displayed output setting in table 8, output to OPR.
(7) Allow measurement to stabilize, then record TI indication in table 8.