TB 9-6625-133-40
(5) Set calibrator to the first Displayed Output value in table A-4, output to OPR.
(6) Allow time for multimeter indication to stabilize then record the value in
(7) Calculate the true applied calibrator current by dividing the multimeter
indication by the test report value (from table A-1) of the resistance standard. Record this
value in the calibrator True Applied column in table A-4.
(8) Repeat technique of (4) through (7) above for remaining rows in table A-4. Use
calibrator +/- key to change polarity.
(9) Set the calibrator output to minimum and disconnect equipment set-up.
Multimeter
Calibrator
Nominal (Ω)
Test Report
Indication
Displayed
True Applied
100 μA
10 k
-100 μA
10 k
1k
1 mA
1k
-1 mA
10
10 mA
10
-10 mA
1
100 mA
1
-100 mA
0.1
1A
0.1
-1 A
10 A1
0.001
0.001
-10 A
Set calibrator OPR/STBY to STBY. Move connection to amplifier and set calibrator OPR/STBY to OPR
1
A-11/(A-12 blank)