Figure A-3. DCI Characterization DCV Zero Connection.

Calibration Procedure For Reference Multimeter, Models 8508A, 8508A/01 And 8508A/02 Page Navigation
  39    40    41    42    43  44
img
. . . . . . . .
TB 9-6625-133-40
(5) Set calibrator to the first Displayed Output value in table A-4, output to OPR.
(6) Allow time for multimeter indication to stabilize then record the value in
"multimeter indication" column of table A-4.
(7) Calculate the true applied calibrator current by dividing the multimeter
indication by the test report value (from table A-1) of the resistance standard. Record this
value in the calibrator True Applied column in table A-4.
(8) Repeat technique of (4) through (7) above for remaining rows in table A-4. Use
calibrator +/- key to change polarity.
(9) Set the calibrator output to minimum and disconnect equipment set-up.
(10) Use values from table A-4 to complete the tables in paragraphs 12 a and 15 e.
Table A-4. DC Current Characterization.
Resistance Standard
Multimeter
Calibrator
Nominal (Ω)
Test Report
Indication
Displayed
True Applied
100 μA
10 k
-100 μA
10 k
1k
1 mA
1k
-1 mA
10
10 mA
10
-10 mA
1
100 mA
1
-100 mA
0.1
1A
0.1
-1 A
10 A1
0.001
0.001
-10 A
Set calibrator OPR/STBY to STBY. Move connection to amplifier and set calibrator OPR/STBY to OPR
1
A-11/(A-12 blank)