TB 9-6625-133-40
CHARACTERIZATIONS
1. Test Report Data
a. Record the test report values for the equipment listed in table A-1.
b. Use values in table A-1 throughout the entire procedure.
NOTE
If there is not an AC/DC difference test report value
(specification) at the frequency of interest (fi) above 1 kHz,
interpolate the specification of interest (Si) between the higher
frequency (f higher) and lower frequency (f lower) using the higher
specification (S higher) and the lower specification (S lower) as
S
S
shown in the formula below.
S
S
+
f
- f
f
-
-f
=
∗
If there is no test report value at the frequency of interest
below 1 kHz, use the test report value at the frequency closest
Test Report
Test Report
Test Report
Test Report
Equipment
Parameter 1
Value 1
Parameter 2
Value 2
0.001 Ω
----------
----------
0.1 Ω
----------
----------
1Ω
----------
----------
10 Ω
----------
----------
100 Ω
----------
----------
1 kΩ
----------
----------
10 kΩ
----------
----------
100 kΩ
----------
----------
1 MΩ
----------
----------
10 MΩ
----------
----------
100 MΩ
----------
----------
1 GΩ
----------
----------
TRANSFER STANDARD
10 V dc
----------
----------
800 Ω
CURRENT SHUNT (1 mA)
----------
400 Hz AC/DC Dif (ppm)
----------
----------
----------
1 kHz AC/DC Dif (ppm)
----------
----------
----------
3 kHz AC/DC Dif (ppm)
----------
----------
----------
5 kHz AC/DC Dif (ppm)
----------
----------
----------
10 kHz AC/DC Dif (ppm)