Non-Volatile Input Offset Compensation Adjustments
Table A-1. Test Report Data.

Calibration Procedure For Reference Multimeter, Models 8508A, 8508A/01 And 8508A/02 Page Navigation
  28    29    30    31    32  33  34    35    36    37    38  
img
. . .
TB 9-6625-133-40
Table 15. Non-Volatile Input Zeros.
Function
Pushbutton sequence
DCV
DCV, INPUT, [ZERO FUNC]
Ω, [2Ω], CONFIG, [FILT] (selected), [4WΩ] (selected), [LoI] (selected)
Ω 4W LoI
INPUT, [ZERO FUNC]
Ω, [2Ω], CONFIG, [FILT] (selected), [4WΩ] (de-selected), [LoI] (selected)
Ω 2W LoI
INPUT, [ZERO FUNC]
Ω, [2Ω], CONFIG, [FILT] (selected), [4WΩ] (selected), [LoI] (de-selected)
Ω 4W NORMAL
INPUT, [ZERO FUNC]
Ω, [2Ω], CONFIG, [FILT] (selected), [4WΩ] (de-selected), [LoI] (de-selected)
Ω 2W NORMAL
INPUT, [ZERO FUNC]
Ω PLUS, [TRU Ω], [2Ω], CONFIG, [LoI] (selected)
Ω Plus Tru LoI
INPUT, [ZERO FUNC]
Ω PLUS, [TRU Ω], [2Ω], CONFIG, [LoI] (de-selected)
Ω Plus Tru Normal
INPUT, [ZERO FUNC]
Ω PLUS, [HIV], [20 MΩ], CONFIG, [FILT] (selected), [4WΩ] (de-selected)
Ω Plus HiV 2W
INPUT, [ZERO FUNC]
Ω PLUS, [HIV], [20 MΩ], CONFIG, [FILT] (selected), [4WΩ] (selected)
Ω Plus HiV 4W
INPUT, [ZERO FUNC]
DCI
DCI, INPUT, [ZERO FUNC]
ACV
ACV, INPUT, [ZERO FUNC]
ACI
ACI, INPUT, [ZERO FUNC]
16. Final Procedure
a. If desired, enter new calibration due date by performing (1) through (4) below.
(1) Set rear panel CALIBRATION switch is set to ENABLE (if necessary).
(2) [Quit].
(3) Keypad entry of date desired.
(4) [ENTER].
b. Ensure that rear panel CALIBRATION switch is set to DISABLE.
c. Deenergize and disconnect all equipment.
d. Annotate and affix DA label/form in accordance with TB 750-25.