TB 9-6625-149-50
(6) Adjust oscilloscope controls to display one positive half cycle of square-wave
pattern. If oscilloscope does not display an optimum square wave with rise time less than
20 nanoseconds, perform b below.
b. Adjustments
(1) Remove left-side cover from TI.
NOTE
In the following adjustments, it will be necessary to replace left
side cover on TI after each adjustment to properly note the
effects on waveform.
(2) Adjust H.F. COMP. R475, C475, L457, and L478 (fig. 4) to obtain waveforms
specified in a(6) above for TI current probe amplifier, types 131-1, -2, and -3. (R)
NOTE
L457 affects a wider portion of the front cover, while L478
affects a narrower portion. Ringing, present in the waveform,
will not exceed 2 percent from nominal top of square wave.
(3) Adjust H.F. COMP. R475, C475, L457, L478 (fig. 4) and C468 (not shown) to
obtain waveform specified in a(6) above for TI current probe amplifier, type 131-4. (R)
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
a. Performance Check
(1) Connect ac/dc voltmeter (A2) between -15-volt test point (fig. 6) and chassis
ground. Ac/dc voltmeter will indicate between -13.5 and -16.5 volts.
indication will remain within limits of (1) above.
b. Adjustments. No adjustments can be made.
22