TB 9-6625-149-50
Table 6. Low Voltage Power Supply Accuracy
Test instrument test point
Ac/dc voltmeter indication (V dc)
Min
Max
+ 100
97
103
+ 225
218.25
231.75
+ 350
339.5
360.5
+ 500
485
515
b. Adjustments.
Repeat a(1) above and
adjust -150 ADJ. R616 (fig. 1)
for a
-150 volt
a. Performance Check.
Connect dc
voltmeter (A5) between HV test point
(fig. 2)
and chassis ground. If dc voltmeter does not indicate between -1310 and -1390 volts,
perform b below.
b. Adjustments. Adjust
HV ADJ. R840 (fig. 1) until dc
voltmeter indicates -1350
volts. (R)
21. Final Procedure
a. Deenergize and disconnect all equipment and replace TI within protective cover.
b. In accordance with TM 38-750, annotate and affix DA Label 80 (U.S. Army
Calibration System). When the TI cannot be adjusted within tolerance, annotate and affix
DA Form 2417 (Unserviceable or Limited Use tag).
SECTION V
PRELIMINARY OPERATIONS FOR CURRENT PROBE AMPLIFIER, TEKTRONIX
TYPES 131-1 THROUGH -4, WITH CURRENT PROBE, TEKTRONIX TYPE P6016.
22. Preliminary Instructions
a. The
instructions outlined in this
section are preparatory to the
calibration process.
Personnel should become familiar with sections I, II, V, and VI of this bulletin before
beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name and item identification number as listed in tables 2 and 3. For the identification of
equipment referenced by item numbers prefixed with A, see table 2, and for prefix B, see
17