Table 15. Audio Generator 1 Output Level.
Audio Frequency Meter

Calibration Procedure For Radio Test Set, Models 2947 And 2947A And Model 2947 Page Navigation
  18    19    20    21    22  23  24    25    26    27    28  
img
. . . . . . . . . .
TB 9-6625-152-24
17. Audio Generator Distortion
a. Performance Check
(1) Connect TI AF GEN OUT through a 10X attenuator to measuring receiver
AUDIO INPUT.
(2) Set TI as listed in (a) through (e) below:
(a) Press AF TEST.
(b) Press AF TEST, [Audio Gen], [Gen1/Gen2] until AF1 FREQ is
highlighted, FREQ, and set to 1 kHz.
(c) Press AF TEST, [Audio Gen], [Gen1/Gen2] until AF1 FREQ is
highlighted, LEVEL, and set to 4 V.
(d) Press AF TEST, [Audio Gen], [Gen1/Gen2] until AF1 FREQ is
highlighted, then press ON OFF key to on.
(e) Press AF TEST, [Audio Gen], [Gen1/Gen2] until AF2 FREQ is
highlighted, then press ON OFF key to off.
(3) Configure measuring receiver for audio distortion measurement. If measuring
receiver indication is not within limits specified in first row of table 17, perform b below.
(4) Repeat technique of (2) (b) through (e) and (3) above for TI frequencies and levels
specified in table 17. If measuring receiver indications are not within limits specified in
table 17, perform b below.
Table 17. Audio Generator 1 Distortion
Test instrument
Measuring receiver
Output
Indication limits
(%)
Frequency (kHz)
Level (V)
1
4
<0.5
1
0.5
<0.5
1
2
4
<1.0
5
4
<1.0
15
4
<1.0
Set measuring receiver low pass filter to 30 kHz.
1
(5) Set TI as listed in (a) through (d) below:
(a) Press AF TEST, [Audio Gen], [Gen1/Gen2] until AF2 FREQ is
highlighted, FREQ, and set to 1 kHz.
(b) Press AF TEST, [Audio Gen], [Gen1/Gen2] until AF2 FREQ is
highlighted, LEVEL, and set to 4 V.
(c) Press AF TEST, [Audio Gen], [Gen1/Gen2] until AF2 FREQ is
highlighted, then press ON OFF key to on.
(d) Press AF TEST, [Audio Gen], [Gen1/Gen2] until AF1 FREQ is
highlighted, then press ON OFF key to off.
(6) Configure measuring receiver for audio distortion measurement with low pass
filter set to none. If measuring receiver indication is not within limits specified in first row
of table 18, perform b below.