Audio Frequency Meter
AF Distortion and Sinad Meter - Continued

Calibration Procedure For Radio Test Set, Models 2947 And 2947A And Model 2947 Page Navigation
  20    21    22    23    24  25  26    27    28    29    30  
img
. . . . . . .
TB 9-6625-152-24
Figure 1. Audio Frequency Meter.
Table 19. Audio Frequency Meter.
Function generator
Test instrument
Output frequency
Indication limits (Hz)
(Hz)
Min
Max
20
19.8
20.2
1k
999.8
1000.2
20 k
19.998 k
20.002 k
(7) Set outputs to minimum and disconnect equipment setup.
b. Adjustments. Perform adjustments in Section IV below.
19. AF Distortion and Sinad Meter
a. Performance Check
(1) Connect equipment as shown in figure 2 below.
(2) Set multimeter for AC voltage measurement in AUTO range.