TB 9-6625-162-24
Test instrument
Calibrator
Output
Range (V)
Min
Max
(V dc)
(V dc)
(V dc)
20, 200, 1000
40, 400, 1000
10
20
40
9.88
10.12
20
20
40
19.78
20.22
35
200
40
34.63
35.37
50
200
400
49.3
50.7
100
200
400
98.8
101.2
200
200
400
197.8
202.2
300
1000
400
295
305
400
1000
400
394
406
500
1000
1000
493
507
600
1000
1000
592
608
800
1000
1000
790
810
1000
1000
1000
988
1012
(4) Set calibrator output to minimum.
b. Adjustments (20/200 Model Only)
(1) Set the TI range to 200 V ac.
(2) Set the calibrator for a 190 V dc output.
(3) Turn VR (2) so that 190.0 is indicated on the LCD.
(4) No other adjustments can be made for DC voltage.
a. Performance Check
NOTE
For the 40/400 auto ranging meter for AC voltage (Defaults to
Lo): Lo: 0-39.99 V (Shifts to Mid at 40.00 V), Mid: 15.0-399.9 V
(Shifts to Lo at 14.9 V and to Hi at 400.0 V), Hi: 150-750 V
(Shifts to Mid at 149 V).
(1) Set TI function switch to AC/Ω and 20V; for the 40/400 auto ranging model set
TI function switch to ~ V. AC should be shown on the upper left corner of the display.
(2) Set calibrator output for the first value in table 4. TI will indicate within limits
specified in first row of table 4. If TI does not indicate within limits specified in table 4,
perform b below.
(3) Repeat technique of (2) above for the remaining settings and indications listed in
table 4. If TI does not indicate within limits specified in table 4, perform b below.