Spurious Responses
Figure 17. Intermodulation distortion products.

Calibration Procedure For Spectrum Analyzer, If, Lf, And Rf Plug-In Units, Models 8552, 8553, 8554, 8555, And 8556 Page Navigation
  55    56    57    58    59  60  61    62  
img
. .
TB 9-6625-1914-24
(2) Position IF plug-in controls as listed in (a) through (h) below:
(a) SCAN TIME PER DIVISION switch to 2 SECONDS.
(b) LOG/LINEAR switch to 10 dB LOG.
(c) LOG REF LEVEL switch to -40 dBm/dBV.
(d) Vernier control fully ccw.
(e) BASE LINE CLIPPER control fully ccw.
(f) VIDEO FILTER switch to 100 Hz.
(g) SCAN MODE switch to INT.
(h) SCAN TRIGGER switch to AUTO.
(3) Connect equipment as shown in figure 16.
Figure 16. Spurious responses - equipment setup.
(4) Adjust function generator no. 2 frequency for 70 kHz at 3 mV.
(5) Adjust signal generator frequency for 90 kHz at amplitude equal to that set in (4) above.