TB 9-6625-194-24
Table 2. Model i200 Ac Current.
Calibrator
TI Applied
Range
applied
calibrator range
current
Min
Max
(A)
(V ac)
(A)
(A ac)
(A ac)
(A ac)
21
1 @ 60 Hz
1
0.470 m
1.530 m
1.05 @ 60 Hz
20
10.5
9.7375 m
11.2625 m
240
2
0.3 @ 60 Hz
20
45
43.60 m
46.40 m
0.48 @ 60 Hz
20
72
70.06 m
73.94 m
1.12 @ 60 Hz
20
168
165.82 m
170.18 m
240
1.52 @ 60 Hz
20
228
0.22522
0.23078
Connect the single turn coil to the Ammeter Calibrator.
1
Connect the 15-turn coil to the Ammeter Calibrator.
2
Table 3. Model i200s Ac Current.
Calibrator
TI Applied
Range
applied
calibrator range
current
Min
Max
(A)
(V ac)
(A)
(A ac)
(V ac)
(V ac)
2
20
1.2 @ 60 Hz
2
18
1.714
1.886
1
0.9 @ 60 Hz
20
9
81.85 m
98.15 m
22
2 @ 60 Hz
30
0.286
0.314
200
0.4 @ 60 Hz
20
60
0.580
0.620
1.2 @ 60 Hz
20
180
1.768
1.832
Connect the single turn coil to the Ammeter Calibrator.
1
Connect the 15-turn coil to the Ammeter Calibrator.
2
Table 4. Model i400 Ac Current.
Calibrator
TI Applied
Range
applied
calibrator range
current
Min
Max
(A)
(V ac)
(A)
(A ac)
(mA ac))
(mA ac)
1
0.8 @ 45 Hz
20
120
117.54
122.46
0.8 @ 400 Hz
20
120
117.54
122.46
1.86 @ 45 Hz
20
279
273.36
284.64
400
1.86 @ 400 Hz
20
279
273.36
284.64
0.24 @ 45 Hz
100
360
352.74
367.26
0.24 @ 400 Hz
100
360
352.74
367.26
Connect the 15-turn coil to the Ammeter Calibrator.
1
Table 5. Model 80i-400 Ac Current.
Calibrator
TI Applied
Range
applied
calibrator range
current
Min
Max
(A)
(V ac)
(A)
(A ac)
(mA ac))
(mA ac)
100 1
375
363.35
386.65
0.25 @ 60 Hz
1.86 @ 60 Hz
20
279
270.23
287.77
400
0.8 @ 60 Hz
20
120
116.0
124.0
0.25 @ 400 Hz
100
375
363.35
386.65
0.25 @ 1 kHz
100
375
359.6
390.4
Connect the 15-turn coil to the Ammeter Calibrator.
1
5