• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 6. Model i-400s Ac Current.
Table 13. Model 80 i-110s Ac Current.

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
TB 9-6625-194-24
Table 9. Model 80i-1000s Ac Current.
Multimeter indication
Calibrator
Ammeter
TI Applied
Range
applied
calibrator range
current
Min
Max
(mV/A)
(V ac)
(A)
(A ac)
(V ac)
(V ac)
20 1
100
1 @ 60 Hz
10
0.96
1.04
1 @ 1 kHz
20
10
0.96
1.04
2
10
0.33 @ 60 Hz
20
49.5
0.4801
0.5099
0.33 @ 400 Hz
20
49.5
0.4801
0.5099
0.33 @ 1 kHz
20
49.5
0.4801
0.5099
0.66@ 60 Hz
20
99
0.9652
1.0148
0.66 @ 1 kHz
20
99
0.9652
1.0148
1
0.33 @ 1 kHz
100
495
0.4851
0.5049
0.6 @ 60 Hz
100
900
0.882
0.918
Connect the single turn coil to the Ammeter Calibrator.
1
Connect the 15-turn coil to the Ammeter Calibrator.
2
Table 10. Model i1000s Ac Current.
Multimeter indication
Calibrator
Ammeter
TI Applied
Range
applied
calibrator range
current
Min
Max
(A)
(V ac)
(A)
(A ac)
(V ac)
(V ac)
1
10
0.9 @ 60 Hz
20
9
0.863
0.937
22
100
2.0 @ 60 Hz
30
0.289
0.311
0.5 @ 60 Hz
20
75
0.730
0.770
0.6 @ 60 Hz
20
90
0.877
0.923
1000
0.6 @ 60 Hz
100
900
0.890
0.910
Connect the 1 turn coil to the Ammeter Calibrator.
1
2 Connect the 15-turn coil to the Ammeter Calibrator.
Table 11. Model i2000 FLEX Ac Current.
Multimeter indication
Calibrator
Ammeter
TI Applied
Range
applied
calibrator range
current
Min
Max
(A)
(V ac)
(A)
(A ac)
(V ac)
(V ac)
1
20
1.2 @ 60 Hz
2
18
17.8 mV
18.2 mV
200
0.4 @ 60 Hz
20
60
0.580
0.620
1 @ 60 Hz
20
150
1.480
1.520
1.3 @ 60 Hz
20
195
1.930
1.970
2000
0.6 @ 60 Hz
100
900
0.880
920
Connect the 15-turn coil to the Ammeter Calibrator.
1
Table 12. Model 80 i-110s Dc Current.
Multimeter indication
Calibrator
Ammeter
TI Applied
Range
applied
calibrator range
current
Min
Max
(mV/A)
(V dc)
(A)
(A ac)
(V dc)
(V dc)
1
100
1
2
1
0.092
0.108
0.3
20
3
0.286
0.314
0.5
20
5
0.480
0.520
0.7
20
7
0.674
0.726
0.9
20
9
0.869
0.932
21
2
10
2
30
0.2875
0.3125
0.46
20
69
0.6067
0.7733
0.6
20
90
0.765
1.035
Connect the single turn coil to the Ammeter Calibrator.
1
Connect the 15-turn coil to the Ammeter Calibrator.
2
7


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.