Table 13. Model 80 i-110s Ac Current.
Final Procedure

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  3    4    5    6    7  8  9    10    11    12    13  
img
TB 9-6625-194-24
Table 15. Model 80 i-1010 Ac Current.
Multimeter indication
Calibrator
Ammeter
TI Applied
Range
applied
calibrator range
current
Min
Max
(mV/A)
(V ac)
(A)
(A ac)
(V ac)
(V ac)
20 1
100
0.63 @ 60 Hz
94.5
0.9161
0.9739
0.43 @ 60 Hz
100
645
0.62565
0.66435
0.63 @ 400 Hz
20
94.5
0.90665
0.98335
0.43 @ 400 Hz
100
645
0.61275
0.67725
Connect the 15-turn coil to the Ammeter Calibrator.
1
Table 16. Model 80 i-410 Dc/Ac Current.
Ammeter
Multimeter indication
TI Applied
Range
Calibrator applied
calibrator
current
Min
Max
range (A)
20 1
400 A
0.8 V dc
120 A dc
0.1115
0.1285
1.8 V dc
20
270 A dc
0.254
0.286
100 mV/A
0.25 V dc
100
375 A dc
0.35375
0.39625
0.25 V ac @ 60 Hz
100
375 A ac
0.35375
0.39625
Connect the 15-turn coil to the Ammeter Calibrator.
1
Table 17. Model i-410 Dc Current.
Multimeter indication
Calibrator
Ammeter
TI Applied
Range
applied
calibrator range
current
Min
Max
(A dc)
(V dc)
(A)
(A dc)
(V dc)
(V dc)
20 1
400
0.8
120
0.1153
0.1247
1.8
20
270
0.26005
0.27995
0.24
100
360
0.3469
0.3731
Connect the15-turn coil to the Ammeter Calibrator.
1
Table 18. Model i-410 Ac Current.
Multimeter indication
TI Applied
Calibrator
Ammeter
Range
current
applied
calibrator range
Min
Max
(A ac)
(A)
(A ac)
(V ac)
(V ac)
(V ac)
20 1
120
0.1153
0.1247
400
0.8 @ 45 Hz
20
120
0.1153
0.1247
0.8 @ 400 Hz
20
270
0.26005
0.27995
1.8 @ 45 Hz
20
270
0.26005
0.27995
1.8 @ 400 Hz
100
360
0.3469
0.3731
0.24 @ 45 Hz
100
360
0.3469
0.3731
0.24 @ 400 Hz
Connect the 15-turn coil to the Ammeter Calibrator.
1
Table 19. Model i-1010 Dc Current.
Multimeter indication
Calibrator
Ammeter
TI Applied
Range
applied
calibrator range
current
Min
Max
(A dc)
(V dc)
(A)
(A dc)
(V dc)
(V dc)
20 1
1000
2.0
300
0.2935
0.3065
0.4
100
600
0.5875
0.6125
0.6
100
900
0.8815
0.9185
Connect the 15-turn coil to the Ammeter Calibrator.
1
Table 20. Model i1010 Ac Current.
Multimeter indication
Calibrator
Ammeter
TI Applied
Range
applied
calibrator range
current
Min
Max
(A ac)
(V ac)
(A)
(A ac)
(V ac)
(V ac)
600
0.36 @ 45 Hz
100
540
0.5287
0.5513
0.36 @ 390 Hz
100
540
0.5287
0.5513
0.26 @ 1 kHz
100
390
0.3778
0.4022
Connect the 15-turn coil to the Ammeter Calibrator.
1
9