Table 10. Model i1000s Ac Current.
Table 15. Model 80 i-1010 Ac Current.

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
.
TB 9-6625-194-24
Table 13. Model 80 i-110s Ac Current.
Multimeter indication
Calibrator
Ammeter
TI Applied
Range
applied
calibrator range
current
Min
Max
(mV/A)
(V ac)
(A)
(A ac)
(V ac)
(V ac)
21
100
1 @ 40 Hz
1
92 m
108 m
1 @ 60 Hz
2
1
92 m
108 m
1 @ 400 Hz
2
1
92 m
108 m
0.3 @ 60 Hz
20
3
0.286
0.314
0.5 @ 60 Hz
20
5
0.480
0.520
0.6 @ 40 Hz
20
6
0.577
0.623
0.6 @ 60 Hz
20
6
0.577
0.623
0.6 @ 400 Hz
20
6
0.577
0.623
0.6 @ 1.5 kHz
20
6
0.559
0.641
0.6 @ 4 kHz
20
6
0.559
0.641
0.6 @ 6 kHz
20
6
0.505
0.695
0.6 @ 10 kHz
20
6
0.505
0.695
10
1 @ 40 Hz
20
10
95.5
104.5
1 @ 60 Hz
20
10
95.5
104.5
1 @ 400 Hz
20
10
95.5
104.5
2
2 @ 40 Hz *
2
30
0.2875
0.3125
2 @ 60 Hz
2
30
0.2875
0.3125
2 @ 400 Hz
2
30
0.2875
0.3125
0.33 @ 40 Hz
20
49.5
0.4351
0.5549
0.33 @ 60 Hz
20
49.5
0.4351
0.5549
0.33 @ 400 Hz
20
49.5
0.4351
0.5549
0.46 @ 40 Hz
20
69
0.6067
0.7733
0.46 @ 60 Hz
20
69
0.6067
0.7733
0.46 @ 400 Hz
20
69
0.6067
0.7733
20 1
1 @ 1.5 kHz
10
92.5
107.5 m
1 @ 4 kHz
20
10
92.5
107.5 m
1 @ 6 kHz
20
10
83.5
116.5 m
1 @ 10 kHz
20
10
83.5
116.5 m
22
1 @ 1.5 kHz *
30
0.2785
0.3215
1 @ 4 kHz
2
30
0.2785
0.3215
1 @ 6 kHz
2
30
0.2515
0.3485
0.33@ 1.5 kHz
20
49.5
0.42025
0.56975
0.33 @ 4 kHz
20
49.5
0.42025
0.56975
0.33 @ 6 kHz
20
49.5
0.3757
0.6143
0.46 @ 1.5 kHz
20
69
0.5860
0.7940
0.46 @ 4 kHz
20
69
0.5860
0.7940
0.46 @ 6 kHz
20
69
0.5239
0.8561
Connect the single turn coil to the Ammeter Calibrator.
1
2 Connect the 15-turn coil to the Ammeter Calibrator.
Table 14. Model 80 i-1010 Dc Current.
Multimeter indication
Calibrator
Ammeter
TI Applied
Range
applied
calibrator range
current
Min
Max
(A dc)
(V dc)
(A)
(A dc)
(V dc)
(V dc)
1
1000
0.5
20
5
3.9
6.1 m
1
20
10
8.8
11.2 m
22
1.3
19.5
0.1811
0.2089
1.0
20
150
0.1455
0.1545
2.0
20
300
0.291
0.309
40.0
100
600
0.582
0.618
0.6
100
900
0.855
0.945
Connect the single turn coil to the Ammeter Calibrator.