SECTION I. IDENTIFICATION AND DESCRIPTION

Calibration Procedure For Signal Standardizer Calibration Fixtures Types 067-0587-01 And 067-0587-02 Page Navigation
1  2    3    4    5    6  
img
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
*TB 9-6625-1960-40
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
SIGNAL STANDARDIZER
CALIBRATION FIXTURES TEKTRONIX,
TYPES 067-0587-01 AND 067-0587-02
Headquarters, Department of the Army, Washington, DC
20 March 2008
Distribution Statement A: Approved for public release; distribution is unlimited.
REPORTING OF ERRORS AND RECOMMENDING IMPROVEMENTS
You can improve this manual. If you find any mistakes or if you know of a way to improve these
procedures, please let us know.  Mail your letter or DA Form 2028 (Recommended Changes to
Publications and Blank Forms) directly to: Commander, U.S. Army Aviation and Missile Command,
ATTN: AMSAM-MMC-MA-NP, Redstone Arsenal, AL 35898-5000. A reply will be furnished to
you.
You may also  send in  your comments electronically to our E-mail address:
2028@redstone.army.mil or by fax 256-842-6546/DSN 788-6546.  For the World Wide Web use:
https://amcom2028.redstone.army.mil. Instructions for sending an electronic 2028 can be found at
the back of this manual.
.
Paragraph
Page
SECTION
I.
IDENTIFICATION AND DESCRIPTION
Test instrument identification ......................
1
2
Forms, records, and reports .........................
2
2
Calibration description ................................
3
2
II.
EQUIPMENT REQUIREMENTS
Equipment required ...................................
4
3
Accessories required ...................................
5
3
III.
CALIBRATION PROCESS FOR
CALIBRATION FIXTURE TEKTRONIX,
TYPE 067-0587-01
Preliminary instructions .............................
6
3
Equipment setup .......................................
7
4
Rep rate ...................................................
8
4
9
5
Gain .........................................................
Risetime ...................................................
10
7
Final procedure .........................................
11
8
SECTION
IV.
CALIBRATION PROCESS FOR
CALIBRATION FIXTURE TEKTRONIX,
TYPE 067-0587-02
Preliminary instructions .............................
12
8
Equipment setup .......................................
13
8
_________
*This bulletin supersedes TB 9-6625-1960-50, dated 21 Apr 1987, including all changes.