TB 9-6625-1960-40
Paragraph
Page
Rep rate ...................................................
Gain ........................................................
Risetime ...................................................
Final procedure .........................................
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Signal Standardizer Calibration Fixtures, Tektronix, Types 067-0587-01 and
067-0587-02. The manufacturer's manuals were used as the prime data sources in
compiling these instructions. The equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.
a. Model Variations.
Variations among models will be noted in text, tables
and figures.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Tektronix, Type 067-0587-01
Repetition rate
Range: 100 Hz to 1 MHz
Accuracy:
20%
Gain
Range: -250 to +250 mV
Accuracy:
0.3%
Risetime
300 ps or less
Tektronix, Type 067-0587-02
Repetition
Range: 10 Hz to 1 MHz
0.1%
Accuracy:
Gain
Range: -250 to +250 mV
0.3%
Accuracy:
Risetime
150 ps or less