Figure 2. Test instrument - right side view (Type 067-0587-01).
Figure 4. Risetime - equipment setup (Type 067-0587-01).

Calibration Procedure For Signal Standardizer Calibration Fixtures Types 067-0587-01 And 067-0587-02 Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . . .
TB 9-6625-1960-40
Figure 3. Gain - equipment setup (Type 067-0587-01).
(5) Press and release stepper pushbutton for nearest zero indication on multimeter
(6) Adjust POSITION control for zero indication on multimeter.
(7) Press and release stepper pushbutton for multimeter indications listed in table 5.
Multimeter will indicate within limits specified.
NOTE
Voltage indications may not be in sequence. Press and release
stepper pushbutton until desired voltage is obtained.
(8) Set oscilloscope mainframe to OFF.
(9) Remove stepper and replace Q118 (fig. 2).
(10) Disconnect multimeter from setup.
Table 5. Gain Voltage Checks
Multimeter indications
Min
Max
(mV)
(mV)
+49.85
+50.15
+99.70
+100.30
+149.55
+150.45
+199.40
+200.60
+249.25
+250.75
-49.85
-50.15
-99.70
-100.30
-149.55
-150.45
-199.40
-200.60
-249.25
-250.75
b. Adjustments. No adjustments can be made.