Figure 6. Test instrument - left side view (Type 067-0587-02).
Final Procedure

Calibration Procedure For Signal Standardizer Calibration Fixtures Types 067-0587-01 And 067-0587-02 Page Navigation
  6    7    8    9    10  11  12  
img
. . . . . . .
TB 9-6625-1960-40
Table 7. Gain Voltage Checks
Multimeter indications
Min
Max
(mV)
(mV)
+49.85
+50.15
+99.70
+100.30
+149.55
+150.45
+199.40
+200.60
+249.25
+250.75
-49.85
-50.15
-99.70
-100.30
-149.55
-150.45
-199.40
-200.60
-249.25
-250.75
NOTE
Voltage indications may not be in sequence. Press and release
stepper until desired voltage is obtained.
(14) Set oscilloscope mainframe switch to OFF.
(15) Remove adapter and stepper from U121 (fig. 5). Replace U115 (fig. 6).
(16) Disconnect multimeter from setup.
b. Adjustments. No adjustments can be made.
16. Risetime
a. Performance Check
(1) Configure oscilloscope to measure risetime.
(2) Set TI TEST switch to VERT or HORIZ + STEP RESP.
(3) Connect equipment as shown in figure 7.
(4) Set oscilloscope and oscilloscope mainframe power switches to ON and allow
approximately 5 minutes for warm-up.
Figure 7. Risetime - equipment setup (Type 067-0587-02).
11