TB 9-6625-1967-35
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
b. Connect autotransformer to
115 V ac
source and
adjust controls for 115 V
output.
c. Pull POWER switch out and allow at least 20 minutes for warm-up.
a. Performance Check
(1) Position controls as listed in (a) through (j) below:
(a) RET LOSS (dB) (SENSITIVITY (DB)) switch and FINE controls to 0 dB.
(b) ZERO REF SET control ccw.
(d) 0-dB SET control fully cw.
(e) DISTANCE dial to 000.
(f) FEET/DIV switch to .5 (METRES/DIV to 1).
(g) Multiplier switch to X10.
(h) Both CABLE DIELECTRIC pushbuttons to out position.
(i) VAR control (DISTANCE CAL controls) fully cw.
(j) IMPLS WIDTH (ns) to 10. (TEST PULSE-SHORT/MED/LONG control to MED.)
(2) Disconnect P21 from VERT AMP/SLOW RAMP BOARD (fig. 1) (located directly
behind center of front panel on left side of board) and connect equipment as shown in figure 1.