Low Frequency Compensation
Figure 4. High Frequency Pulse Response.

Calibration Procedure For Digitizing Oscilloscope, Agilent, Models 54600A, 54601A And 54602A Page Navigation
  12    13    14    15    16  17  18  
img
. . . .
TB 9-6625-197-24
Figure 3. Low Frequency Compensation.
17. High Frequency Pulse Response
a. Connect pulse generator to TI channel 1.
b. Set pulse generator for a 10 ns, 200 mV pulse output.
c. Press TI Autoscale key.
d. Set TI timebase to 10 ns/div.
e. Adjust TI Volt/Div for approximately 6 divisions of deflection displayed.
f. Adjust CH 1 high frequency response adjustment (fig. 4) for 1.5 minor division of
overshoot (6%).
g. Repeat a through f above for remaining channels.
h. Set all outputs to minimum and disconnect equipment setup.
i. Perform paragraphs 8 through 12 above.
18