SECTION IV. ADJUSTMENT PROCESS
Low Frequency Compensation

Calibration Procedure For Digitizing Oscilloscope, Agilent, Models 54600A, 54601A And 54602A Page Navigation
  10    11    12    13    14  15  16    17    18  
img
. . .
TB 9-6625-197-24
Figure 2. Power Supply Voltages.
15. Self-Calibration
a. Connect TI DC CALIBRATOR (rear panel) to multimeter input.
b. Press TI keys as listed in (1) through (3) below:
(1) Print/Utility.
(2) [Self-Test Menu].
(3) [DAC].
c. Configure multimeter for DC voltage measurement.
d. Multimeter will indicate between -0.0005 and +0.0005 V dc.
e. Disconnect multimeter.
f. Set TI rear panel CALIBRATION switch to UNPROTECTED (up position).
g. Press TI Print/Utility key, then press [Self Cal Menu].
h. Press TI [Load Defaults].
i. When TI displays "Default calibration factors loaded", press [Vertical].