TB 9-6625-1980-24
(11) Press POWER PUSH OFF and replace assembly A5A1 in its original position.
b. Adjustments (Fig. 2)
(1) Adjust A4A8R9 +25 V ADJ for multimeter indication of 25.0 V (R).
(2) Adjust A5A1R5 -25 ADJ for multimeter indication of -25.0 V (R)
(3) Adjust A5A1R11 -5 V ADJ for multimeter indication of -5.0 V (R).
16. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.
14