• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Output Voltage and Linearity
Output Voltage and Linearity - Continued

Calibration Procedure For Dc Voltage/Current Calibrator Model 3330B Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. .
TB 9-6625-1980-24
Figure 2. Test instrument - top interior view.
Table 3. Output Voltage and Linearity
Test instrument
Multimeter indication (V dc)
dial settings
Min
Max
9.999999
9.998799
10.001199
8.888888
8.887821
8.889955
7.777777
7.776844
7.778710
6.666666
6.665866
6.667466
5.555555
5.554888
5.556222
4.444444
4.443911
4.444977
3.333333
3.332933
3.333733
2.222222
2.221955
2.222489
1.111111
1.110978
1.111244
1.000000
0.999880
1.000120
2.000000
1.999760
2.000240
4.000000
3.999520
4.000480
8.000000
7.999040
8.000960