TB 9-6625-1980-24
a. Performance Check
(1) Disconnect all equipment from TI except autotransformer.
(2) Press POWER PUSH OFF pushbutton. If TI meter does not indicate 0 (zero),
adjust to 0 (zero) by turning adjusting screw located below meter face.
(3) Press pushbuttons as listed in (a) through (d) below:
(a) METER READING VOLTAGE.
(b) MODE VOLTS and 100-1000.
(c) POWER PUSH, pushbutton to ON.
(d) POWER OPR.
(4) Adjust TI decade dials until VOLTS-MA meter indicates 1000 V. If decade dials
do not indicate between 990.0000 and 1010.0000, perform b (1) and (5) below.
(5) Repeat technique of (4) above at VOLTS-MA meter indications listed in table 6.
If decade dials do not indicate within limits specified, perform b (3) below.
VOLTS-MA meter
Decade dial indications
indications (V)
Min
Max
790.0000
810.0000
800
590.0000
610.0000
600
390.0000
410.0000
400
190.0000
210.0000
200
(6) Press POWER STDBY RESET pushbutton.
(7) Press pushbuttons as listed in (a) through (d) below:
(a) METER READING CURRENT.
(b) MODE MA and 100-1000.
(c) CROWBAR ON.
(d) POWER OPR.
(8) Adjust TI decade dials until VOLTS-MA meter indicates 100 mA. If decade
dials do not indicate between 99.00000 and 101.00000, perform b (4) and (5) below.
(9) Press POWER STDBY/RESET and CROWBAR OFF pushbuttons.
b. Adjustments
(1) Set decade dials to indicate 1000.0000.
(2) Adjust A4A4R5 METER ADJ (fig. 2) for VOLTS-MA meter indications of 1000 V (R).
(3) Readjust A4A4R5 METER ADJ for best overall in-tolerance condition.
(4) Set decade dials to indicate 100.00000.