TB 9-6625-1980-24
(27) Adjust R18 100 V CAL (fig. 2) for 100.00000 V indication on multimeter (R).
(28) Set TI decade dials to indicate 700.00000.
(29) Adjust R21 1000 V CAL (fig. 2) for 700.00000 V indication on multimeter (R).
a. Performance Check
(1) Connect equipment as shown in figure 1.
(2) Position controls as indicated in (a) through (d) below:
(a) METER READING CURRENT pushbutton pressed.
(b) MODE MA and 1-10 pushbuttons pressed.
(c) Decade dials to indicate .10000000 (first decade to 10).
(d) POWER OPR pushbutton pressed.
(3) Measure voltage across dc current shunt.
between -0.100060 and -0.099940 perform b (1) below.
(4) Press POWER STDBY/RESET pushbutton.
(5) Repeat technique of (2) through (4) above at TI settings using shunt settings
listed in table 5. Multimeter will indicate within limits specified.
Test instrument
indications (V dc)
MODE MA
Shunt
Decade dial
settings
settings
settings
Min
Max
1-10
.5000000
1 mA
-0.050030
-0.049970
100 A
1-10
.1000000
-0.100060
-0.099940
1
-0.0999401
10-100
10.000000
10 mA
-0.100060
10-100
5.000000
10 mA
-0.050030
-0.049970
10-100
1.000000
1 mA
-0.100060
-0.099940
2
-0.0999402
100-1000
100.00000
100 mA
-0.100060
100-1000
50.00000
100 mA
-0.050030
-0.049970
100-1000
10.00000
10 mA
-0.100060
-0.099940
1If
not within tolerance, perform b (2) below.
2If
not within tolerance, perform b (3) below.
b. Adjustments
(1) Adjust R1, 1 MA CAL (fig. 2) for a -.100000 V dc indication on multimeter (R).
(2) Adjust R5 10 MA CAL (fig. 2) for a -.100000 V dc indication on multimeter (R).
(3) Adjust R9 100 MA CAL (fig. 2) for a -.100000 V dc indication on multimeter (R)