Figure 3. Adjustment location - types DC503A and DC503AOPT01.
Time Base

Calibration Procedure For Counters Types Dc 501, Dc 501Opt01, Dc 503, Dc 503Opt01, Dc 503A, Dc 503Aopt01, Dc 504, Dc 504A, Dc 508, Dc 508Opt01, Dc 508Aopt01 And Dc 509 Page Navigation
  9    10    11    12    13  14  15    16    17    18    19  
img
. . . . . . .
TB 9-6625-1982-24
(5) Repeat technique of (4) above at 1 kHz, 100 kHz, 1 MHz, and 5 MHz.
Function/arbitrary generator amplitude will indicate 20 mV rms or less at each frequency.
(6) Disconnect function/arbitrary generator from TI and connect signal generator
RF OUTPUT to TI CH 1 INPUT, using adapter, cable, and termination.
(7) Set TIMING switch to 100 mS (change as needed) and repeat technique of (4)
above at 10 MHz, 50 MHz, 100 MHz, and 125 M Hz.  Signal generator will indicate
20 mV rms or less at 10 MHz and 50 MHz and 35 mV rms or less at 100 MHz and 125 MHz.
b. Adjustments. No adjustments can be made.
22. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
SECTION VI
CALIBRATION PROCESS FOR TEKTRONIX, TYPE DC504
23. Preliminary Instructions
a. The instructions outlined in paragraphs 23 and 24 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
24. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance checks where applicable.
a. Equipment Setup. Position TI controls as listed in (1) through (4) below:
DISPLAY TIME FUNCTION switch to 1 s, change as needed.
(1)
(2)
SOURCE INT EXT switch to EXT.
(3)
TRIGGER LEVEL control to midrange.