TB 9-6625-199-24
(4) Repeat technique of (2) and (3) above for remaining ranges and applied values
listed in Range and Applied columns. If TI indication is not within limits specified, perform
b below.
(5) If applicable, disconnect equipment setup.
b. Adjustments. Refer to TI manufacturer's manual for corrective action.
a. Performance Check
(1) Connect TI DC CURRENT terminals to calibrator OUTPUT terminals,
observing polarity.
NOTE
For current ≥2.2 A, use amplifier CURRENT OUTPUT
terminals and set BOOST On when required.
(2) Set TI controls for the first range listed in the Range column. Some TIs require
the test leads be changed to different jacks for range changes.
(3) Set calibrator for the first value listed in the Applied column and set OPR/STBY
switch to OPR. If TI indication is not within limits specified, perform b below.
(4) Repeat technique of (2) and (3) above for remaining ranges and applied values
listed in Range and Applied columns. If TI indication is not within limits specified, perform
b below.
(5) If applicable, disconnect equipment setup.
b. Adjustments. Refer to TI manufacturer's manual for corrective action.
a. Performance Check
(1) Connect TI AC CURRENT terminals to calibrator OUTPUT terminals,
observing polarity.
NOTE
For current ≥2.2 A, use amplifier CURRENT OUTPUT
terminals and set BOOST On when required.
(2) Set TI controls for the first range listed in the Range column. Some TIs require
the test leads be changed to different jacks for range changes.
(3) Set calibrator for the first value listed in the Applied column and set OPR/STBY
switch to OPR. If TI indication is not within limits specified, perform b below.
(4) Repeat technique of (2) and (3) above for remaining ranges and applied values
listed in Range and Applied columns. If TI indication is not within limits specified, perform
b below.
(5) If applicable, disconnect equipment setup.
b. Adjustments. Refer to TI manufacturer's manual for corrective action.