TB 9-6625-199-24
(3) Set function generator for the first value listed in the Applied column. If TI
applied input level is not specified, adjust the function generator output for a stable TI
display. If TI indication is not within limits specified, perform b below.
(4) Repeat technique of (2) and (3) above for remaining ranges and applied values
listed in Range and Applied columns. If TI indication is not within limits specified, perform
b below.
(5) If applicable, disconnect equipment setup.
b. Adjustments. Refer to TI manufacturer's manual for corrective action.
a. Performance Check
(1) Connect test leads and adapters required to TI Input and position them so they
are not touching.
(2) Set TI controls as necessary to measure the first value listed in the Applied
column and zero TI (press TI REL switch) if required.
listed in the Applied column. If TI indication is not within limits specified, perform b
below.
(4) Repeat technique of (2) and (3) above for remaining ranges and applied values
listed in Range and Applied columns. If TI indication is not within limits specified, perform
b below.
(5) If applicable, disconnect equipment setup.
b. Adjustments. Refer to TI manufacturer's manual for corrective action.
a. Performance Check
CAUTION
Do not turn off cool/heat source at temperatures higher than
100 C. This may create a hazardous situation. Select a set-
point lower than 100 C and allow the instrument to cool before
turning off.
NOTE
Perform this check if temperature probe is supplied with TI. If
not TI will receive a calibration label (DA 163) stating that
Temperature not calibrated.