TB 9-6625-1995-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Dual Time Base, Tektronix, Type 7B92A. The manufacturer's manual was
used as the prime data source in compiling these instructions. The equipment being
calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Model variations are described in the text.
b. Time and Technique. The time required for this calibration is approximately 1
hour, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
parameters
Performance specifications
Triggering
Range: At least 3.5 V
Horizontal timing
Delayed sweep:
Range: 0.2 s/div to 0.5 ns/div
Accuracy:1
2%
0.2 s/div to 20 ns/div
3%
10 ns/div and 5 ns/div
4%
2 ns/div and 1 ns/div
5%
0.5 ns/div
Delaying sweep:
Range: 0.2 s/div to 10 ns/div
Accuracy:1
2%
0.2 s/div to 20 ns/div
3%
10 ns/div
See footnote at end of table