TB 9-6625-1995-24
Table 1. Calibration Description - Continued
Test instrument
parameters
Performance specifications
Range: 0 to 9.8 times the TIME/DIV OR DLY TIME switch setting
Differential delay
from 0.2 s/div to 10 ns/div
time measurement
Range: TIME/DIV OR DLY TIME switch setting 0.2s/div to 0.1 s/div
accuracy
Accuracy1 Within 0.75% of measurement +0.25% of FS 2 (both DLY
TIME MULT dial settings at 0.05 or greater)
within 0.75% of measurement +0.5% of FS 2 +5 ns (one or
both DELAY TIME MULT dial settings < 0.05)
Range: TIME/DIV OR DLY TIME switch setting 50 ns/div to 10 ns/div
Accuracy: 1 Within 1.0% of measurement +0.5% of FS 2
(both delay times equal to or greater than 25 ns)
within 1.0% of measurement +1% of FS 2 +5 ns
(one or both delay times < 25 ns)
1Accuracy applies over the center 8 divisions.
2FS is 10 times the TIME/DIV OR DLY TIME switch setting.
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The accessories listed in table 3 are issued as indicated in
paragraph 4 above and are used in this calibration procedure. When necessary, these items
may be substituted by equivalent items, unless specifically prohibited.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Output capability: 8 V p-p, 50 kHz
Fluke, Model 5720A
(5700A/EP) (p/o MIS-35947)
Fluke, Model 8840A/AF05
Measurement capability:1 V dc
(AN/GSM-64D)
Fluke, Model 5820A-5C-GHZ
Time markers:
(5820A-5C-GHZ)
CALIBRATOR
Range: .2 s/div to 2 ns/div
Accuracy: 0.5%
1Use recorded indications.
.