TB 9-6625-2023-35
(2) Connect calibrator SENSE HI and LO to TI Ω SIGNAL (4 WIRE) HI and
LO.
(3) Set controls as listed in (a) through (g) below:
(a) FUNCTION 4 WIRE kΩ key to on.
(b) RANGE 1 key to on.
(c) HIGH RESOLN key to on.
(d) TRIGGER INTERNAL key to on.
(e) MATH OFF key to on.
(f) AUTO CAL key to on.
(g) GUARD pushbutton pressed to on.
(4) Set calibrator for a 1 kΩ output, EX SNS on, and 2 wire Comp off then, using
output adjustment controls, set calibrator control display Reading equal to TI
indication. If calibrator control display Error indication is not within 0.0046%, perform
b below.
(5) Press RANGE 100 key to on.
(6) Set calibrator for a 100 kΩ output then, using output adjustment controls, set
calibrator control display Reading equal to TI indication. If calibrator control display
Error indication is not within 0.0047%, perform b below.
(7) Repeat technique of (5) and (6) above using settings and indications in table 4.
Test instrument
Calibrator
Control
display Error
RANGE
HIGH RESOLN
Output
indications
(%)
settings
settings
settings
100 kΩ
100
off
0.007
1 kΩ
1
off
0.006
100 Ω
.1
off
0.011
10 kΩ
10
on
0.0071
10 kΩ
10
off
0.009
1 MΩ
1k
off
0.019
1 MΩ
1k
on
0.0146
10 MΩ
10 k
on
0.1006
10 MΩ
10 k
off
0.105
(8) Set calibrator for a 100 Ω output, EX SNS off, and 2 wire Comp on.
(9) Press TI FUNCTION 2 WIRE kΩ key to on and repeat technique of (5) and
(6) above using settings and indications in table 5.
8