TB 9-6625-2023-35
(29) Set calibrator for a 100 V, 40 kHz output. Adjust A15C34 (fig. 1) until TI
indicates 100.010 V (R).
(30) Set calibrator to STANDBY.
(31) Repeat (18) through (30) above until optimum adjustment is obtained.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
12