TB 9-6625-2049-24
Test instrument
parameters
Performance specifications
Harmonic distortion1
(relative to fundamental at
Accuracy: No harmonic greater than -65 dB
full output)
Fundamental frequency: 50 to 200 kHz
Accuracy: No harmonic greater than -60 dB
Fundamental frequency: 200 kHz to 2 MHz (200 kHz to 1 MHz
for option 002)
Accuracy: No harmonic greater than -40 dB
Fundamental frequency: 2 to 15 MHz
Accuracy: No harmonic greater than -30 dB
Fundamental frequency: 15 to 20 MHz
Accuracy: No harmonic greater than -25 dB
-30 dB at 80% modulation, 10 kHz, and 0 V dc
Amplitude modulation
AM distortion:
offset
20 ns, ( 125 ns for option 002) at full
Square wave
Rise time and fall time:
output
.02% of period +3 ns
Symmetry:
5% of p-p amplitude at full output (<10% of peak
Overshoot:
amplitude for option 002)
3 s retrace time, positive or negative ramps
Ramp retrace
Accuracy: 5 x 10-6 of selected value
Frequency
1 Hz to 20 999 999.999 Hz
Sine wave
Range:
(option 002: .02 Hz to 25.6 kHz)
1 Hz to 10 999 999.999 Hz
Square wave
Range:
(option 002: .02 Hz to 25.6 kHz)
1 Hz to 10 999 999.999 Hz
Triangle
Range:
(option 002: .02 Hz to 10 kHz)
1 Hz to 10 999 999.999 Hz
Positive slope ramp
Range:
(option 002: .02 Hz to 10 kHz)
1 Hz to 10 999 999.999 Hz
Negative slope ramp
Range:
(option 002: .02 Hz to 10 kHz)
Variable 719.9 with respect to arbitrary
Phase offset
Range:
starting phase, or assigned zero phase
A
0.2
Accuracy:
Tolerance relative to
mplitude
programmed amplitude
0.1 dB
2
Amplitude accuracy with no
Sine wave: .001 Hz to 100 kHz
1.0%
Square wave: .001 Hz to 100 kHz
1.5%
Triangle: .001 Hz to 2 kHz
into 50 load (no dc offset )
5%
2 kHz to 10 kHz
1.5%
Ramps: .001 Hz to 500 Hz
10%
500 Hz to 10 kHz
Tolerance relative to
programmed amplitude at 1 kHz
(attenuator range 1) into a 50 load
0.3 dB
Sine wave: 100 kHz to 20 MHz
10%
Square wave: 100 kHz to 10 MHz
See footnotes at end of table.
3