TB 9-6625-2049-24
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in the
calibration procedure. The following peculiar accessory is also required for this calibration:
50 feedthrough termination, Hewlett-Packard, Model 11048C.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
AUDIO ANALYZER
Output: 4.5 V, 10 kHz
Boonton, Model 1121
(1121)
Range: 99,9950 s to 60,000,300 Hz
Fluke, Model PM6681/656
FREQUENCY COUNTER
(PM6681/656)
Accuracy: 1.25 x 10-8
Capability: Time interval A to B
MEASURING RECEIVER
Measuring Receiver system
N5530S consisting of:
Volts: .683 to .732 V rms
Accuracy: 0.85%
Spectrum Analyzer Agilent,
Model E4440A (E4440A),
AM: 0 to 80%
Power Meter Agilent, Model
AM distortion: -30 dB at 10 kHz Distortion:
E4419B (E4419B), and
-60 dB at 50 Hz
Sensor Module Agilent,
Model N5532A opt 504 (504)
Range: -20.225 to +20.225 V dc
Hewlett-Packard, Model
Accuracy: 0.05%
3458A (3458A)
Range: 0.3416 to 20.4 V ac
100 Hz to 100 kHz
Accuracy: 0.3%
Frequency: 1 kHz to 10 MHz
Agilent, OS-303/G (OS-
303/G)
Amplitude: 1.8 to 2.2 V p-p
Accuracy: 2.5%
Capabilities
Duty cycle: 49.7 to 50.3%
Overshoot: 5%
Rise time and fall time: <20 ns
Biddle Gray, Model 71-631
Range: 470
(7910328)
SPECTRUM ANALYZER
Frequency: 50 kHz to 20 MHz
(AN/USM-677)
Input: -20 dBm and 40 V
Display capability: <-65 dBc
SYNTHESIZER/LEVEL
Frequency: 0.1 MHz
Hewlett-Packard, Model
Amplitude: 13 dBm
3335AOPT 001-K06
(MIS-35938)
Datum, Model ET6000-75
TIME/FREQUENCY
Accuracy: 5 x 10-8
(13589305)
WORKSTATION