Figure 1. Test instrument - left view.
Table 4. Magnified Sweep Timing - Continued

Calibration Procedure For Delaying Time Base Type 7B85 Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . . . .
TB 9-6625-2055-24
Table 3. Sweep Timing
Test instrument
TIME/DIV
Oscilloscope calibrator
switch
err indication
settings
MARKER output
(%)
20
ns
20
ns
2.5
s
s
1.5
0.1
0.1
s
s
1.5
0.2
0.2
s
s
1.5
0.5
0.5
s
s
1.5
1
1
s
s
1.5
2
2
s
s
1.5
5
5
s
s
1.5
20
20
s
s
1.5
50
50
0.1
ms
0.1
ms
1.5
0.2
ms
0.2
ms
1.5
0.5
ms
0.5
ms
1.5
1
ms
1
ms
1.5
2
ms
2
ms
1.5
5
ms
5
ms
1.5
20
ms
20
ms
1.5
50
ms
50
ms
1.5
0.1
s
0.1
s
1.5
0.2
s
0.2
s
1.5
0.5
s
0.5
s
1.5
1
s
1
s
4
2
s
2
s
4
5
s
5
s
4
(10) Press and release MAG pushbutton to out (X10) position.
(11) Repeat technique of (3) and (4) above, and set oscilloscope calibrator MARKER
output and TI TIME/DIV switch settings as listed in table 4. At each setting, oscilloscope
calibrator will display one time marker per division between second and 10th vertical
graticule line within percent of error listed in table 4.
NOTE
Due to bandwidth limitation of vertical amplifier, TI magnified
10, 20, and 50 ns switch positions are omitted from table 4.
Table 4. Magnified Sweep Timing
Test instrument
Oscilloscope calibrator
TIME/DIV
switch
err indication
settings
MARKER output
(%)
s
10
ns
2.5
0.1
s
20
ns
2.5
0.2
7