TB 9-6625-2059-24
Table 1. Calibration Description Continued.
Test instrument parameters
Performance specifications
Power monitor:
Range: 1 to 1000 MHz
Frequency
0 to 4, 0 to 40, and 0 to 400 W
Power
Accuracy: 7% of reading. 3% FS from 1 to 600 MHz
20% of reading 3% FS from 600 to 1000 MHz
10 and 100 mV/div; 1 and 10 V/div
External vertical input
Horizontal sweep
Accuracy: 10%
10 and 1 ms/div; 100 and 10 s/div
Frequency:
Variable tone
Range: 10 to 20 kHz
Accuracy: 0.01%
Range: 1 kHz
Fixed tone
Accuracy: 20 Hz
Output level
0 to 2.5 V rms minimum either tone into 150 Ω
Distortion
<1.5% 10 to 100 Hz
<0.7%, 100 Hz to 20 kHz
<2%, 1 kHz fixed tone
Range: 70 dB from -30 to -100 dBm
Dynamic range
Accuracy: 4 dB relative to -50 dBm
Continuous from 0.5 to 5 MHz from center frequency
(1 to 10 MHz span)
Frequency error meter:
Sensitivity
2.0 V above 1 MHz
Range: 1.5 kHz, 5 kHz, 15 kHz FS
Accuracy: 0.5 kHz on 15 kHz range
0.2 kHz on 5 kHz range
0.1 kHz on 1.5 kHz range
Deviation kHz meter
Range: 2, 6, and 20 kHz ranges
Accuracy: 4% reading 3% FS
Range: 3.2, 8, 150, 600 Ω and HI-Z (1 MΩ)
Accuracy: 10%
Ac load
Ac volts
Range: 0.1 to 300 V in 8 ranges
Accuracy: 5% FS
Dc volts
Range: 0.1 to 300 V in 8 ranges
Accuracy: 3% FS
Distortion
Range: 0 to 10% and 0 to 30%
Accuracy: 1.5% on 0 to 10%
3.5% on 0 to 30
Range: 0 to 20 dB
Sinad
Accuracy: 1.5 dB
Range: 0 to 100%
Am %
Accuracy: 10%
Range: 0 to 100 MΩ
Accuracy: 5% at midscale
1 Frequency
not verified below 1 MHz.
2 Output
level below -80 dBm is indirectly verified.
3