TB 9-6625-2059-24
(20) BFO-RF LEVEL control fully cw.
(21) HI LVL-Vx100-NORM switch to HI LVL.
(22) AUTO-ZERO-OFF-BATT switch to AUTO ZERO.
(23) PWR-OFF-BATT switch to OFF.
b. If FREQ ERROR, DEVIATION, and MM-100E meters do not indicate zero, adjust
to zero with adjustment screw located below each meter face.
c. Connect TI to autotransformer. Connect autotransformer to a 115 V ac source and
adjust for 115 V output.
d. Set PWR OFF BATT switch to PWR and allow at least 15 minutes for warm-up.
a. Performance Check
(1) Connect frequency counter to TI 10 MHz REF OUT.
(2) Adjust REF CAL (front panel) until frequency counter indicates between
9,999,995 and 10,000,005 Hz. If indication cannot be obtained, perform b (1) and (2) below.
(3) Connect TRANS-RCVR output to frequency counter.
(4) Set FREQUENCY MHz switches to the first setting in table 4.
Frequency
counter will indicate within specifications listed in table 4.
(5) Repeat technique of (4) above for settings and indications listed in table 4.
Test instrument
Frequency counter indications
(Hz)
FREQUENCY MHz
switch settings
Min
Max
111 111 1
111,111,044
111,111,156
222 222 2
222,222,088
222,222,311
333 333 3
333,333,133
333,333,467
444 444 4
444,444,177
444,444,622
555 555 5
555,555,222
555,555,778
666 666 6
666,666,266
666,666,933
777 777 7
777,777,311
777,778,089
888 888 8
888,888,355
888,889,244
999 999 9
999,999,400
1,000,000,400
010 000 0
9,999,995
10,000,005
050 000 0
49,999,975
50,000,025
090 000 0
89,999,955
90,000,045
001 000 0
999,995
1,000,005
005 000 0
4,999,995
5,000,005
009 000 0
8,999,995
9,000,005