Table 4. Frequency Accuracy Check.
Table 5. Modulation Frequency Accuracy.

Calibration Procedure For Radio Test Set, An/Grm-114A (Ifr, Model Fm/Am-1100S) Page Navigation
  3    4    5    6    7  8  9    10    11    12    13  
img
. .
TB 9-6625-2059-24
b. Adjustments
(1) Adjust REF CAL (front panel) to its mechanical center.
(2) Adjust TCXO (fig. 1) until frequency counter indicates 10,000,000 Hz.
Figure 1. Test Instrument - Top View.
8