TB 9-6625-2059-24
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets, AN/GSM-287 and AN/GSM-440, or Secondary Reference Set, NSN
4931-00-621-7878. Alternate items may be used by the calibrating activity. The items
selected must be verified to perform satisfactorily prior to use and must bear evidence of
current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provided a four-to-one ratio between the
standard and TI. Where the four-to-one ratio cannot be met, the actual accuracy of the
equipment is shown in parentheses.
5. Accessories Required. The accessories listed in table 3 are issued as indicated in
paragraph 4 above and are used in this calibration procedure. When necessary, these items
may be substituted by equivalent items, unless specifically prohibited.
Table 2. Minimum Specifications of Equipment Required.
Manufacturer and model
Common name
Minimum use specifications
(part number)
ATTENUATOR
Hewlett-Packard, Model 355D
(355D)
Weinschel, Model 9918-20dB
ATTENUATOR (FIXED)
(9918-20) (9918 Set)
Accuracy: 0.625 dB
AUDIO ANALYZER
Capability: 0.7 to 50% from
Boonton, Model 1121
20 to 9999.9 Hz
(1121)
Range:
105 to 125 V ac
Ridge, Model 9020A
Accuracy: 1%
(9020A),
Fluke, Model 5720A (5720A) (p/o
CALIBRATOR
Range:
0.095 to 315 V ac
Accuracy: 1.25%
MIS-35947); w/ amplifier, Fluke
Range:
0.097 to 309 V dc
5725A/AR (5725A/AR)
Accuracy: 0.75%
Range: 1 MHz to 1 GHz
Fluke, Model PM6681/656
FREQUENCY COUNTER
Accuracy:0.0000125%
(PM6681/656)
Range: 999.9 to 20002 Hz
Accuracy:0.0025%
Range: 1.8 k Hz to 900 MHz
Measuring receiver system N5531S
MEASURING RECEIVER
Accuracy:1.75%
consisting of: Spectrum Analyzer,
Measurement range: 0 to 90 dBm
Agilent Model E4440A (E4440A),
Accuracy:0.625 dB
Power meter, Agilent Model E4419B
(E4419B), or Power meter, Agilent
Range: 10 to -10 dBm at 10 MHz
Model N1911A (N1911A) and Sensor
to 1 GHz
modules, Agilent Model N5532A opt.
Accuracy:2.5%
504 (504), 518 (518), 526 (526)
Range: 0 to 16.75 V dc
Hewlett-Packard, Model 3458A
Accuracy:0.16%
(3458A)
Range: 0.054 to 3.5 V Pac
Accuracy:1.25%
Range: 0 to 600 Ω
Accuracy:2.5%