TB 9-6625-2059-24
14. Oscilloscope Deviation and Frequency Error
a. Performance Check
(1) Position controls as listed in (a) through (d) below:
(a) GEN-RCVR switch to RCVR.
(b) FREQUENCY MHz switches to all zeroes. If, necessary, adjust ZERO
RCVR (front panel) adjust for 0 indication on FREQ ERROR meter.
(c) DEV-VERT V/DIV outer switch to 15 kHz.
(d) VERT control to center oscilloscope trace on horizontal center graticule line.
between -9.5 and -10.5 kHz, perform b (1) below.
NOTE
Trace will appear approximately 2 major divisions below
horizontal centerline.
(3) If FREQ ERROR meter does not indicate negative between 9.5 and 10.5 kHz
perform b (2) below.
(4) Position controls as listed in (a) through (d) below:
(a) FREQUENCY MHz switches to all zeroes.
(b) EXT DEV-VERT V/DIV outer switch to 6 kHz.
(c) FREQ ERROR meter switch to 5.
(d) VERT control to center oscilloscope trace on horizontal center graticule line.
between -3.8 and -4.2 kHz, perform b (3) below. If FREQ ERROR meter does not indicate
negative 3.8 and 4.2 kHz, perform b (4) below.
(6) Position controls as listed in (a) through (d) below:
(b) DEV-VERT V/DIV outer switch to 1.5 kHz.
(c) FREQ ERROR meter switch to 1.5.
(d) VERT position control to center oscilloscope trace on horizontal center
graticule line.
between -0.9 and -1.1 kHz, perform b (3) below. If FREQ ERROR meter does not indicate
negative between 0.9 and 1.1 kHz, perform b (5) below.
(8) Set FREQUENCY MHz switches to 1255000.
(10) Adjust signal generator frequency for 125.500 MHz and amplitude for -70 dBm.
(11) Record frequency error indication of FREQ ERROR meter.