TB 9-6625-2059-24
(6) Configure oscilloscope calibrator for level sine output.
(7) Set DEV-VERT V/DIV outer switch to .1.
(8) Adjust oscilloscope calibrator frequency to 50 kHz and amplitude for 5 divisions
of vertical deflection on TI display.
(9) Increase oscilloscope calibrator frequency to 1 MHz. If the displayed amplitude
is less than 3.5 major divisions, perform b (2) below.
b. Adjustments
(1) Adjust HORIZ GAIN A4A4R57 (fig. 1) for 1 marker per major division.
(2) Adjust A4A4C1 (fig. 1) for displayed amplitude greater than 3.5 major divisions.
a. Performance Check
(1) Set oscilloscope calibrator output to off and remove 50 Ω termination. Reconnect
to TI SCOPE IN.
(2) Set DEV-VERT V/DIV outer switch to 1, SWEEP outer control to 1 mS and
inner control to CAL (detent).
(3) Configure oscilloscope calibrator for a voltage function, 1 kHz, 5 V output.
(4) Adjust oscilloscope calibrator edit knob control for 5 divisions of vertical
deflection on TI oscilloscope. If oscilloscope calibrator error display readout does not
indicate within 10%, perform b below.
(5) Repeat technique of (3) and (4) above at switch settings listed in table 14. If
error display readout does not indicate within 10%, perform b below while adjusting for
best in-tolerance condition on all ranges.
Test instrument
Oscilloscope volts switch settings
DEV-VERT V/DIV switch settings (V)
0.01
50
mV
0.1
0.5
V
10
50
V
b. Adjustments
(1) Adjust oscilloscope calibrator edit knob control until error display readout
indicates 0.00.
(2) Adjust VERT GAIN A4A4R7 (fig. 1) until TI oscilloscope indicates 5 divisions of
vertical deflection.