TB 9-6625-2071-35
Test instrument
Signal generator
indications
Amplitude (≤)
Frequencies
30
MHz
25 mV1
3.00000
60
MHz
50 mV
6.00000
90
MHz
50 mV
9.00000
100
MHz
50 mV
10.00000
15 mV for AN/USM-459 and 5328AF096.
1
(11) Disconnect signal generator from TI.
(12) Set TI FREQ RESOLUTION, N switch to 1Hz, 106 and CHANNEL B DC/AC
switch to AC.
(13) Repeat (3) through (11) above substituting 20 Hz for 10 Hz and 0.000002 for
b. Adjustments
(1) Connect signal generator OUTPUT RF to TI INPUT B using 50 Ω feedthrough
termination (omit termination for option H60).
(2) Set signal generator for a 40 MHz, 25 mV output.
NOTE
Do not adjust sensitivity below 10 mV.
(3) Adjust A19Rl8 (fig. 1) (A12R28 (fig. 1) for option 041) for a stable TI 4.00000
indication (R).
(4) Slowly decrease signal generator output and repeat (3) above.
(5) Connect TI CHANNEL B MARKER OUTPUT to oscilloscope Vertical 1 input
using a 50 Ω feedthrough termination.
(6) Set oscilloscope channel 1 for DC input coupling, 1MΩ input, 500mV/div
amplitude and 10 s/div sweep.
(7) Set oscilloscope triggering for positive slope, auto level, channel 1 source
and DC coupling.
(8) Connect function generator Function Outputs Unbalanced to TI INPUT B
using 50 Ω feedthrough termination.
(9) Set function generator for a sine wave, 10 kHz, 42 mVp-p, 50 Ω output.
(10) Adjust oscilloscope controls as necessary to view waveform. Adjust A12R28
(fig. 1) for a
symmetrical square wave on
oscilloscope (R).
(11) Adjust A12R54 (fig. 1) for a stable TI 0.00100 indication (R).
11. Channel C
Sensitivity (Option 030, AN/USM-459, and
5328AF096)
a. Performance Check
(1) Connect signal generator OUTPUT RF to TI CHANNEL C input.